Shop

MX165CT / GS-3576S

This 3.5 MHU CT tube is specifically designed as a replacement for use with GE HiSpeed* Advantage and GE CT/i* CT systems.
Weight 85 kg
Brand

Varex

Material

Steel

MHU

3.5 MHU

Description

Product Description The GS-3576S is a 6.5” (165 mm) 140 kV, 2.6 MJ (3.5 MHU) maximum anode heat content, rotating anode insert. This insert is specifically designed for CT Scanners. The insert features a 7° tungsten-rhenium facing on molybdenum with a graphite backed target and is available with the following nominal focal spot: 0.9 x 0.5 1.2 x 1.2 IEC 60336 Loading Factor for slit focal: Small - 75 kV, 100 mA Large - 75 kV, 200 mA Maximum Anode Cooling Rate: 7,100 W (10 kHU/sec) Maximum continuous anode heat dissipation: 9,695 W (820 kHU/min) Nominal CT Anode Input Power: Small - 28 kW IEC 60613:2010 Large - 42 kW IEC 60613:2010 Nominal CT Scan Power Index: Small - 27.6 kW IEC 60613:2010 Large - 42 kW IEC 60613:2010 Reference Axis: Perpendicular to port face. This insert is intended for use in the MX165NP housing. All trademarks property of the respective manufacturer.